← Tools for your practice

Tool · Protocol

The Piper TMJ MRI Scan Protocol

Three tiers: Screening, Basic, and Full. Each higher tier includes every sequence in the tier below it. Give the protocol to the imaging center with every order.

The PDF opens in a new tab. Close that tab to return to this page.

The three tiers

Tier 1

Screening

4 sequences

Determines whether a derangement is present. Start here.

  • Axial T2
  • Sagittal STIR (Bite 1, FSCP)
  • Coronal T2 (Bite 1, FSCP)
  • Sagittal Proton Density (Bite 2, IEP)

Tier 2

Basic

7 sequences

Stages the joint for treatment planning. The Screening four, plus:

  • Sagittal T1 (Bite 1, FSCP)
  • Sagittal Proton Density (Prop Open, TCP)
  • Sagittal Proton Density (Closed Mouth, MIP)

Tier 3

Full

12 sequences

For surgical planning and medicolegal documentation. The Basic seven, plus:

  • Axial T1 Localizer
  • Coronal T2 Localizer (no occlusion)
  • Coronal T1 (Bite 1, FSCP)
  • Coronal Proton Density (Bite 2, IEP)
  • Coronal Proton Density (Prop Open, TCP)

Every tier uses bilateral surface coils, scanning the left joint and then the right, in the axial, sagittal, and coronal planes.

The four occlusal positions

FSCP

Fully seated condylar position

Requires a bite registration.

MIP

Maximal intercuspation position

The habitual bite. Practice and verify it with the patient before scanning.

IEP

Incisal edge position

Requires a bite registration.

TCP

Translated condylar position

Registered with a ratchet prop during the MR exam.

Read the Piper Classification of TMJ disorders →